Biochemistry Analyzer B800
Biochemistry Analyzer B800 Immuno assay system is based on the principle of magnetic particle, glow Chemiluminescence technology…. It is meant for the diagnosis of Thyroids, Fertility,Cardiac,Prenatal screening, Glyco metabolism, Bone metabolism,ToRCH,Tumormarkers, Anaemia, Inflammation monitoring,Kidney function,Infectious like HIV,HBsAG,HCV & Syphilis..etc. It is a fully automated Random access analyser with true stat facility for sample. The throughput is up to 100 Tests/hour. 100 samples with continuous loading facility and 24 different reagents can be loaded at a time. Reagent Area is provided with cooling facility.Liquid level and clot detection facility available for sample/reagent probes. Reagent kits are Liquid stable and barcoded for identification . Calibrators and Internal quality controls are provided. Samples identification through Bar code scanner. System can connect with LIS. Auto dilution facility with preferable ratios available. Real time calibration curve used for assay calibration. Wide menu with special parameters available like TB IGRA,AMH,Prenatal,17-OH progesterone, HS CRP, Aldosterone, Angiotensin , Plasma Renin Activity , Hepatic Fibrosis , Reverse T3 etc.
Specification
| Item | Details |
|---|---|
| General Performance | |
| Throughput | Constant 800 photometric tests/hour; 400 tests/hour for assays with 3 or 4 reagents; ISE: maximum 600 tests/hour. |
| Principle | Colorimetry, turbidimetry, ISE. |
| Analysis Methods | Endpoint method, rate method, two‑point method; ISE method (optional). |
| Scalability | Integrated with chemiluminescent analyzer; capability to be connected into TLA. |
| Sample System | |
| Sample Types | Whole blood, serum, plasma, urine, pleural effusion, cerebrospinal fluid. |
| Sample Probe | Automatic liquid level detection; vertical & horizontal collision protection; clot detection; bubble detection; sample auto‑dilution. |
| Sample Volume | 1.5–35 µL in 0.1 µL steps. |
| Reagent System | |
| Reagent Probes | 2 reagent probes, for R1(R4) and R2(R3) respectively; automatic liquid level detection; vertical & horizontal collision protection; bubble detection. |
| Reagent Tray | R1(R4) 45 positions; R2(R3) 45 positions. |
| Reagent Volume | 15–350 µL in 0.5 µL steps. |
| Reagent Refrigeration | 2–8 °C; 24‑hour uninterrupted refrigeration. |
| Reaction System | |
| Reaction Cuvettes | 160 reusable cuvettes; 8‑step auto‑washing with vertical collision protection function for acid, alkaline, deionized water cleaning and wipe dry. |
| Reaction Volume | 100 µL–360 µL. |
| Reaction Temperature | Water bath, 37 ± 0.1 °C. |
| Optical System | |
| Light Source | Halogen‑tungsten lamp; service life not less than 2000 hours. |
| Photometer | Flat‑field holographic concave grating. |
| Wavelengths | 16 wavelengths. |
| Working Conditions | |
| Dimension | L × W × H: 1570 × 952 × 1205 mm. |
| Weight | < 400 kg. |
| Water Quality / Pressure / Flow | Water quality: resistivity not less than 1 MΩ/CM; inlet water pressure: 100–392 kPa; inlet water flow: not less than 50 L/H. |
| Power Supply | 220 ± 22 V, 50 ± 1 Hz, 1500 VA. |