Microbiology MALDI-TOFMS1000

4.5

Specification

Autobio Microbiology Solution Specifications
Accessories & Ready-to-Use Reagents
Cat No. Description Specification
MSA01 Sample Pretreatment Reagent 80T
MSA02 Calibrator 80T
MSA03 CHCA Matrix 0.5mL × 4 Bottles
TL00X3/0001 Target Slide 10 pcs
TL00X3/0002 Target Slide Holder 2 pcs
Complete Microbiology Solution
  • Culture: Microbiological Culturing Plate, Automated Blood Culture System
  • Identification: Autof MS 1000, Drying Chemoenzymatic Culturing Medium
  • AST: Mycoplasma IES, Fungi ID/AST, Auto MIC Strip
Sample Workflow
  1. Pick an isolated colony
  2. Smear on the target slide and add matrix
  3. Load to Autof MS 1000 system
  4. Software generates spectrum, compares with reference database to give identification
  5. Final review by microbiologist
Hardware
  • Innovative optical design enhances resolution
  • PCA chip improves accuracy and stability of ion focusing
  • High vacuum turbo molecular pump ensures stable vacuum environment
  • Unique design of signal detection system improves sensitivity
  • Automatic laser energy adjustment reduces manual operation
Software Acquisition System:
  • Efficient data processing system saves laser cost
  • Rapid identification mode reduces analysis time
  • Real-time spectrum preview available
Analysis System:
  • Customized report templates and database self-build
  • Multiple functional analysis
  • Internet multi-user operation supported
Database
  • Adopts reference or quality-controlled strains from various sources
  • Massive database expansion capability to meet needs of multiple routines
Laser 337 nm nitrogen laser, 400 μJ, 60 Hz
Analyzer Linear flight tube of 1.05 m, resolution ≥ 600, mass range 2,000–20,000 Da
Mass Range 2,000–20,000 Da
Dimensions 440 × 720 × 380 mm
Weight Approx. 60 kg
Power Supply AC 100–240 V, 50/60 Hz, 800 VA
Installation Requirements Temperature 15–30°C, Humidity < 70%, Vibration-free